Florida Department of Agriculture and Consumer Services

10/07/2025 | Press release | Distributed by Public on 10/07/2025 16:28

Recall - TRUE METRIX (Blood Glucose Meters) produced by Trividia Health, Inc. - Potential for defective LCD displays that can affect product performance

Summary

Company Announcement Date: October 07, 2025 FDA Publish Date: October 07, 2025 Product Type: Medical Devices Reason for Announcement:
Recall Reason Description
Potential for defective LCD displays that can affect product performance
Company Name: Trividia Health, Inc. Brand Name:
Brand Name(s)
TRUE METRIX
Product Description:
Product Description
Blood Glucose Meters

Company Announcement

(FT. LAUDERDALE, FL) -October 7, 2025 - Trividia Health, Inc., announced today that it is initiating a voluntary recall of a limited number of TRUE METRIX® Self-Monitoring Blood glucose meters co-branded under the Good Neighbor Pharmacy® brand and distributed in the United States.

The company has determined that 601 TRUE METRIX® Meters, lot number KD0746, manufactured September 4, 2025 and distributed September 8, 2025, through September 16, 2025, may have the potential for defective LCD displays that can affect product performance.

It is possible that the LCD display for the affected products may show partial or missing numerical segments or characters or show ghosting (fading) of numerical segments or characters. As a result, it is possible that users could misinterpret a test result or experience a delay in obtaining test results. For users with low glucose (hypoglycemia), this could result in a delay in treatment or therapy decisions.

Trividia Health has not received any reports of patient injuries related to this voluntary recall.

Consumers who believe they have affected product can confirm in one of the following ways:

  • Call Trividia Health Customer Care Department toll-free at 1-888-835-2723 Monday-Friday 8AM-8PM EST (excluding holidays), our representatives will assist in checking if you have affected Product by looking up your meter's serial number
  • Contact us via email at [email protected] so we can connect with you to walk you through how you check your meter's serial number located on the back of the TRUE METRIX® meter to determine if it is one of the effected meters in lot KD0746
  • By checking the lot number (KD0746) printed on the side of the Product's box
  • Visit https://www.trividiahealth.com/productnotice for a quick search using the meter's serial number.

If you have an affected Product, the Trividia Health Customer Care Department will help with return and replacement information.

Consumers may continue to use the TRUE METRIX® Self-Monitoring Blood Glucose System under the Good Neighbor Pharmacy® brand for lots not included in this voluntary recall.

The company is sending notifications to its customers, including pharmacies, mail order companies and distributors where the TRUE METRIX® meters are sold in the United States.

Patient safety is our top priority at Trividia Health. This voluntary recall is being conducted in coordination with the U.S. Food and Drug Administration FDA to quickly resolve this matter.

About Trividia Health

Trividia Health, Inc., is a global health and wellness company based in Fort Lauderdale, Florida and a leading developer, manufacturer and marketer of advanced performance products for people with diabetes. With products sold under TRUE and store brand labels, the company is the exclusive partner and supplier of affordable, high-quality blood glucose monitoring and health and wellness solutions for the world's leading retail pharmacies, distributors and mail service providers. For more information, please visit: https://www.TrividiaHealth.com.

Company Contact Information

Consumers: Trividia Health Customer Care Department 1-888-835-2723
Florida Department of Agriculture and Consumer Services published this content on October 07, 2025, and is solely responsible for the information contained herein. Distributed via Public Technologies (PUBT), unedited and unaltered, on October 07, 2025 at 22:28 UTC. If you believe the information included in the content is inaccurate or outdated and requires editing or removal, please contact us at [email protected]